Technical Specifications
General R Series Specs
Size: 8.2 in (20.8 cm) high x 10.5 in (26.7 cm) wide x 12.5 in (31.7 cm) deep.
Weight: 13.6 lbs (6.2 kg) with OneStep™ Cable and SurePower™ battery pack; 15.2 lbs (6.9 kg) with paddles.
Power Sources: AC Mains: 100 to 120 V AC (50/60 Hz), 220 to 240 V AC (50 Hz); Zoll R Series Battery: Rechargeable lithium-ion battery pack.
Low Battery Indicator: A “LOW BATTERY” message appears on the monitor when there are less than 15 minutes of ECG monitoring.
R Series Design Standards: Meets or exceeds applicable requirements of UL 60601, AAMI DF80, IEC 60601-2-4, EN 60601-2-25, and 60601-2-27.
Patient Safety: All patient connections are electrically isolated.
Environmental: Zoll R Series Defibrillator Operating Temperature: 0°C to 40°C; Storage and Shipping Temperature: -20°C to 60°C; Humidity: 5% to 95% relative humidity, noncondensing; Vibration: IEC 68-2-6 and IEC 68-2-34; Shock: IEC 68-2-27, 50 g 6mS half-sine; Operating Pressure: 594 to 1060 millibars; Particle and Water Ingress: IEC 529, IP22; Electromagnetic
Compatibility (EMC): CISPR 11 Class B Radiated and Conducted Emissions; Electromagnetic Immunity: AAMI DF80, EN 61000-4-3 to 10 V/m; Electrostatic Discharge: AAMI DF80, EN 61000-4-2; Conducted Susceptibility: EN 61000-4-4, 61000-4-5, 61000-4-6.
R Series Defibrillator
Waveform: Rectilinear Biphasic.™
Patient Impedance Range: 15 to 300 ohms.
Energy Selections: 1 to 10, 15, 20, 30, 50, 75, 100, 120, 150, and 200 joules selected using controls on the front of the defibrillator or sternum paddles. (Note: When using appropriate pediatric resuscitation electrodes, the 75-joule setting is replaced by 70- and 85-joule settings.)
Smart Step Energy Levels: Zoll R Series Defibrillator automatically escalates energy through a configured adult or pediatric protocol.
Energy Display: Shown on the R Series monitor for both selected and delivered.
Charge Time: Less than 7 seconds with a new, fully charged Zoll R Series battery (first 15 charges to 200 joules); longer charge times may result in a depleted or older R Series battery.
Synchronized Mode: Synchronizes R Series Defibrillator pulse to patient’s R wave. The “SYNC” message is displayed on the monitor, and markers are shown on both the monitor and the recorded ECG.
Charge Controls: Control from the front of the defibrillator or apex paddle.
Paddles: External apex/sternum paddles; adult plates slide off to expose the pediatric electrode surface.
Code Readiness Testing: Verifies R Series Defibrillator hardware, therapy delivery cable (with both paddles and electrodes), electrode condition, and expiration (with select OneStep electrodes) without needing a separate test fixture.
ECG Monitoring
Patient Connection: 3-lead ECG cable, 5-lead ECG cable, paddles, or hands-free electrodes; selectable by front panel switch.
Input Protection: Fully defibrillator protected. Circuits designed to prevent distortion of ECG signal by pacer pulse.
Implanted Pacemaker Spike Display: R Series Circuits are designed to detect most implanted pacemaker spikes and display a marker on the ECG trace.
Bandwidth: 0.5 to 21 Hz (-3dB) standard; 0.05 to 150 Hz diagnostic with configurable options of 0.5 Hz to 40 Hz or 1 Hz to 21 Hz.
Lead Selection: I, II, III aVR, aVL, aVF, V, P1, P2, P3 with OneStep Pacing electrode.
ECG Size: 0.5, 1.0, 1.5, 2.0, or 3.0 cm/mV display on the monitor.
Heart Rate Display: 0 to 300 bpm ±5%.
Heart Rate Alarm: User selectable for tachycardia at 60 to 280 bpm; for bradycardia at 20 to 100 bpm. On/Off status is displayed on the R Series Defibrillator screen.
CPR Dashboard Featuring Real CPR Help
They are activated when OneStep Complete, OneStep CPR, or OneStep Pediatric CPR electrodes are connected.
Detection Technology: Accelerometer.
Zoll R Series Compression Depth: Detected between 0.75 in (1.9cm) and 3.0 in (7.6 cm), with an accuracy of ±0.25 inches (0.6 cm).
R Series Compression Rate: Detected between 50 and 150 compressions per minute.
Release Bar: Ensures proper release off of the chest.
Feedback: Configurable audio and visual prompts for rate and depth are issued when compressions fall outside AHA/ERC recommendations.
CPR Idle Time Display: Indicates elapsed time since last detected chest compression.
Perfusion Performance Indicator (PPI): Integrates compression depth and rate to rapidly visualize CPR performance per AHA/ERC recommendations.
See-Thru CPR Filter
The Zoll R Series Defibrillator removes the compression-related artifacts from the ECG via an adaptive filtering technique.
R Series Display
Screen Type: Color, VGA liquid crystal display (LCD).
Screen Size: 6.5 inches (16.5 cm) diagonally.
Sweep Speed: 25 mm/sec.
Viewing Time: 5 seconds with standard display format.
Channels: 3.
Information: Heart Rate, Leads/Pads, Alarm On/Off, Selected Energy, Delivered Energy, User Prompts and Warnings, Code Readiness Test Results, SpO2, NIBP, EtCO 2, Pacer Functions, Code Markers, CPR Dashboard.
Zoll R Series Defibrillator Battery Packs
Type: 10.8 V (nominal) rechargeable lithium-ion.
Capacity: 5.8 amp hours.
Weight: 1.7 lb (0.77 kg).
Recharge Time: 5 hours or less with the integral charger.
Operating Time: >4 hours of continuous ECG monitoring; 100 maximal energy (200 joules) discharges; 3.5 hours of continuous ECG monitoring and pacing at 60 mA, 80 ppm.
Recorder
Technology: 90 mm thermal array; 80 mm grid width.
Speed: 25 mm/sec, 6-second delay.
Printing Modes: Manual or automatic.
Annotations: Time, date, defibrillation energy, patient impedance, heart rate, pacer output, QRS synchronization marker, ECG size, ECG lead, alarm, defibrillator test results, analyze ECG, ECG bandwidth. I/O, Storage, Communications
Sync In: 0 to 5 V (TTL Level) pulse, active high, 5 to 15 in duration, no closer than 200 ms apart; energy transfer begins within 25 ms of the leading edge of the sync-in pulse.
Marker Out: 0 to 5 V (TTL Level) pulse, active high, 10 ms in duration, the leading edge of the pulse occurs within 35 ms of the R wave peak.
ECG Output: 1.0 V/cm of deflection on a recorder; <25 ms delay from the patient ECG input. Card Slot: Compact flash compatible.
Internal Memory: Disk on the chip.
Advisory Defibrillation
R Series Shock Advisory Function: Evaluates ECG rhythm to determine if shock delivery is required.
Shockable Rhythms: Ventricular fibrillation with amplitudes >100 μV and wide-complex ventricular tachycardia with rates >150 bpm for adults or >200 bpm for pediatric applications. Refer to the Operator’s Manual for details on sensitivity and specificity performance.
Zoll R Series Protocol Configurations: Configurable for either CPR or shock-first-driven protocols. Energy sequences can be configured for single or multiple shocks with fixed or escalating energy levels. The CPR interval length is configurable in 1-minute increments up to 4 minutes.
External Pacing (not included in all configurations)
Type: VVI demand; asynchronous (fixed rate) when used without ECG leads or in asynchronous (ASYNC) pacing mode.
Pulse: Rectilinear, constant current: 40 ms ±2 ms; variable 0 to 140 mA ±5% or 5 mA, whichever is greater. The rate is variable from 30 to 180 ppm ±1.5%.
Output Protection: Fully defibrillator protected and isolated.
OneStep Pacing: Eliminates the need to connect separate ECG leads when used in conjunction with OneStep Pacing and OneStep Complete electrodes.
Pulse Oximetry with Masimo SET® Technology (not included in all configurations)
Saturation Range: 1-100% (%SpO2) with a resolution of 1%.
Pulse Rate Range: 25-240 ppm with a resolution of 1 ppm.
Saturation Accuracy: Non-motion conditions ±2% for adults/pediatrics; ±3% for neonates. During motion ±3% for all patients.
Pulse Rate Accuracy: Non-motion conditions ±3 ppm. During motion ±5 ppm.
Mainstream CO2 Capnostat 5 Sensor (not included in all configurations)
Principle of Operation: Nondispersive infrared (NDIR) single-beam optics, dual-wavelength, no moving parts.
Warm-up Time: Full specifications within 2 minutes at an ambient temperature of 25°C. Capnogram in 20 seconds.
Environmental: Operating Temperature: 0°C to 45°C, Storage and Shipping Temperature: -40°C to 70°C.
NIBP (not included in all configurations)
Patient Population: Adult, Pediatric.
Method: Oscillometric.
Control: Automatic and manual measurement.
WiFi Capable
WiFi 802.11 a/b/g/n Ambicom-specific 1100C-CF Card
P/N 8005-000101-01 compatibility.
Typical Readiness File: 750K.
Typical Code Data File: 1.2 MB